Column Dead Volume Formula:
From: | To: |
The column dead volume (Vdead) is the volume of mobile phase contained within a chromatography column that is not occupied by the stationary phase. It's important for determining retention times and system dwell volumes.
The calculator uses the dead volume formula:
Where:
Explanation: The formula calculates the empty space in the column by multiplying the column volume (πr²L) by the porosity factor (ε).
Details: Knowing the dead volume is crucial for method development in chromatography, as it helps determine void time, system dwell volume, and retention factor calculations.
Tips: Enter column radius and length in centimeters, and porosity (typically 0.6-0.8 for packed columns). All values must be positive numbers.
Q1: What's the difference between dead volume and void volume?
A: Dead volume refers to the entire system's extra-column volume, while void volume specifically refers to the column's empty space.
Q2: How does particle size affect dead volume?
A: Smaller particles typically have lower porosity, slightly reducing dead volume, but the main effect is on peak shape and resolution.
Q3: What porosity value should I use?
A: For packed columns, use 0.6-0.8. For monolithic columns, consult manufacturer specifications as porosity can be higher.
Q4: How accurate is this calculation?
A: It provides a theoretical estimate. Actual dead volume may vary due to packing efficiency and column hardware.
Q5: Can I use this for capillary columns?
A: Yes, but ensure you're using the correct inner diameter measurement and account for any wall coatings.